Envoyez-nous un email :
info@labcompanion.cn-

-

Demander un appel :
+86 18688888286
I. Core Function of Refrigerant Oil: Ensuring Long-term Stable Operation of the Compressor
As the "heart" of the refrigeration system of the high and low temperature rapid temperature change chamber, the compressor's operating status directly determines the overall performance of the equipment. Refrigerant oil is the key medium to maintain the normal operation of the compressor, mainly undertaking four core functions: lubrication, cooling, sealing, and cleaning. Specifically, refrigerant oil can effectively reduce the friction loss of moving parts inside the compressor, lower the operating temperature during operation, seal the refrigerant leakage channel, and clean internal carbon deposits and impurities, thereby fundamentally extending the service life of the compressor. Lab Companion high and low temperature rapid temperature change chambers are equipped with high-quality semi-hermetic piston compressors, matched with specially adapted refrigerant oil. The purity of the oil and the matching degree of its performance parameters with the equipment directly affect the operating efficiency, energy consumption level and overall service life of the refrigeration system. Therefore, strictly following the refrigerant oil replacement specifications and controlling the replacement cycle is a core link in the daily maintenance of the equipment and a key measure to avoid refrigeration system failures.
II. Lab Companion Refrigerant Oil Replacement Specifications: Detailed Cycle Standards and Operation Steps
Combined with industry standards and the structural characteristics of Lab Companion equipment, clear cycle standards and operation specifications have been formulated for the replacement of refrigerant oil in semi-hermetic piston compressors to ensure that the replacement process is scientific and standardized, and the equipment performance is not affected.
(I) Replacement Cycle
1. First Replacement: The first refrigerant oil replacement must be completed within 500 hours of equipment operation. The purpose is to remove metal debris and impurities generated during the factory inspection and initial operation of the equipment to avoid wear of internal parts of the compressor;
2. Routine Replacement: After the first replacement, the subsequent replacement cycle should be flexibly adjusted according to the oil quality and actual operating conditions. Under normal operating conditions, it should be replaced every 10,000 to 12,000 hours;
3. Emergency Replacement: If abnormal oil quality (such as discoloration, impurities, peculiar smell, etc.) is found during daily inspection, there is no need to wait for the fixed cycle. The equipment must be shut down immediately for inspection and refrigerant oil replacement to avoid the expansion of faults.
(II) Replacement Operation Steps
Lab Companion provides users with a standardized operation guide with clear and operable steps, as follows:
1. Preprocessing: Start the equipment and run it until the refrigeration system reaches the normal working temperature, and continue to run for 1 hour to ensure that the refrigerant oil circulates fully, facilitating the discharge of internal impurities;
2. First Oil Drainage: Turn off the equipment power supply. After the system pressure drops to the normal range, drain all the old oil in the oil tank, cooler and all filters to ensure no residual old oil;
3. Secondary Flushing: Add new refrigerant oil to the oil tank to the minimum oil volume required for equipment operation (about 1/2 of the oil tank volume), start the machine and run it to the normal working temperature for 1 hour, then shut down and drain the flushing oil in the oil tank again;
4. Tertiary Flushing and Cleaning: Repeat the above secondary flushing steps. After completing the third oil drainage, thoroughly clean the filters and filter screens to remove residual impurities and old oil traces;
5. Adding New Oil: Add Lab Companion specially adapted refrigerant oil to the oil tank until it reaches the normal oil level specified by the equipment, ensuring sufficient oil volume and no leakage;
6. Test Run: Restart the equipment and run it for 15 minutes. Closely observe whether the oil level is normal, whether there is abnormal noise during equipment operation, and whether the refrigeration system is stable. Only after confirming that there is no abnormality can it be put into normal use.
III. Oil Quality Judgment Skills: Timely Avoid Equipment Operation Hidden Dangers
In addition to strictly following the fixed replacement cycle, the oil quality status can be quickly judged by observing the oil quality, color and smell of the refrigerant oil during daily operation and maintenance, so as to avoid compressor damage and refrigeration system failures caused by oil deterioration in advance. The specific judgment standards are as follows:
1. Normal State: The refrigerant oil is transparent and clear, with no visible impurities or peculiar smell, indicating that the oil quality is good and can continue to be used normally;
2. Mild Abnormality (Yellow Oil Color): If the oil color turns light yellow or yellow, carefully observe whether there are impurities or burning smell, and check whether air enters the refrigeration system causing oil oxidation; if no moisture or impurities are detected, it can be temporarily not replaced, but the inspection cycle should be shortened; if the oil color is dark and accompanied by slight impurities, the new oil should be replaced in time, and the system should be simply cleaned;
3. Severe Abnormality (Brown Oil Color): If the oil color turns dark brown or black brown and is accompanied by obvious burning smell, stop the machine immediately and detect the resistance value of the compressor motor winding; if the winding insulation is good, the refrigerant oil must be completely replaced and the refrigeration system must be fully cleaned; if the winding insulation is abnormal, contact after-sales service to check the compressor failure to avoid further damage to the equipment.
IV. Lab Companion After-sales Support: Full-cycle Escort to Reduce Maintenance Costs
To help users complete the refrigerant oil replacement standardizedly and avoid operational risks, Lab Companion provides full-cycle after-sales support services to comprehensively reduce users' maintenance costs and extend the service life of the equipment.
1. Standardized Guidance: Provide users with a detailed refrigerant oil replacement operation manual and video guidance to ensure that users can quickly grasp the operation points;
2. Professional Guidance Service: Equipped with a professional after-sales team. On-site maintenance services are available in China to guide the refrigerant oil replacement steps and explain oil quality judgment skills on site to solve users' operational problems; On-site services are not provided overseas, but one-on-one online guidance services are available to answer various questions during the operation process and ensure the standardized completion of the replacement work;
3. Specially Adapted Consumables: Provide Lab Companion equipment specially adapted refrigerant oil to ensure that the oil quality and performance are fully matched with the equipment, and eliminate equipment failures caused by the use of inferior refrigerant oil or non-adapted refrigerant oil;
4. Full-cycle Operation and Maintenance Support: Regularly remind users to replace refrigerant oil, provide equipment maintenance suggestions, help users reduce compressor wear, improve refrigeration efficiency, reduce energy consumption, and keep the equipment in a stable operating state for a long time, providing reliable support for various testing work.